TECHNICAL SPECIFICATIONS OF THE LANDSTATLANDSTATPotentiostat Yes Galvanostat Yes Impedancemeter Yes Maximum current ±100 mA Compliance voltage ±17.5 V Max potential applied ±5 V Voltage range ±1 V, ±2 V and ±5 V Potential accuracy < 0.1 % FSR (Full Scale Range) Potential resolution 0,003 %. Best: 30 µV Maximum scan rate 200 V/s Current ranges 9 (12 with low current probe) With standard board ±1 nA to ±100 mA With low current option 1 pA to 10 nA Current accuracy < 0.1 % FSR (Full Scale Range) Current resolution 0.003 % FSR. Best: 30 fA Potentiostat rise/fall time < 2 µs Input impedance > 1 TΩ (//20 pF) Interfaces USB 2.0 Acquisition time >100 µs IR Compensation Manual and automatic feedback Electrodes Connections 2, 3 and 4 A/D Converter 16 bits EIS capability 10 µHz to 1 MHz. Up to 5 MHz with impedance option Analog I/O Yes, 1 External current booster 5 A to 20 A Flotting option No, versatile connectivity Analog filters Yes, 1 µs to 1 s Digitals filters Yes Dimensions D x W x H in cm: 37 x 58 x 25 Power requirement 90-264 Vac 47-63 Hz, 30 VA Weight 15.5 kg PC Software Cell cable lenght Sur demande Temperature control -10°C to 105°C 14°F to 221°F Sous réserve de modifications sans préavis |